Nova-Tech Digi–Sense 12-Channel Benchtop Scanning Thermometer User Manual
Page 14

13
Front panel displays and controls
1 STORE button
1 STORE button
Push to store one record cycle of the current set of readings in memory.
(Same as LOG, except only one record is stored.)
When a numerical entry is required and the NUM icon is on, push to
enter a ‘1’.
The word “STORE” a and the record number being stored b appear in
the display for about 3 seconds. Then the display returns to normal.
“STORE” c appears, and remains in the display when one or more record
cycles are in storage (STORE and/or LOG).
Use to save a particular data record.
•
A record includes readings from all 12 channels plus the time and
date.
•
Up to 4,680 records may be stored. Memory is shared with LOG.
When no more data can be stored the display reads “STORE
FULL”.
•
To clear (erase) stored records, push CLEAR, then push STORE.
“STORE” disappears from the display.
+ -
A L A R M
E X I T
M E N U
1
S T O R E
2
L O G
3
R E C A L L
4
P R I N T
5
M A X
6
M I N
7
A V G
8
H O L D
9
S C A N
°
F
0
C L E A R
.
1
4
°
F
STORE
c
1
STORE
°
F
STORE
a
b
BA 13-25
10/12/01, 11:03 AM