Nova-Tech Digi–Sense 12-Channel Benchtop Scanning Thermometer User Manual

Page 36

Advertising
background image

35

Front panel displays and controls continued

How to adjust the scan rate

NUM

How to adjust the scan rate
The scan rate is the interval between the start of individual scans of all
twelve channels.

The default setting is 00:04 (4 seconds).
It takes 4 seconds to do one 12 channel scan, so a scan rate of 00:04
results in continuous scanning.
With the scan rate at 00:20 there will be 3 scans per minute.
The maximum scan rate that can be set is one hour (60:00).

+ -

A L A R M

E X I T

MENU

1

S T O R E

2

L O G

3

R E C A L L

4

P R I N T

5

M A X

6

M I N

7

A V G

8

H O L D

9

S C A N

NUM

1b

1, 3b

3a

1a

2

1c

0

CLEAR

To adjust the scan rate:

1. Push MENU until SCAN is displayed or push MENU at least

once, then push SCAN.
SCAN 1a is displayed and the current setting 1b flashes.
NUM 1c appears and flashes indicating number keys 2 are
active.

2. Push number keys 2 to enter the scan rate. (For example, 36:12

is entered by pushing 3 - 6 - 1 - 2.)

3a. Push EXIT to save the setting and return to scanning, or

3b. Push MENU to switch to the next setup topic (adjust the print

rate).

BA 32-36