Front panel displays and controls continued

7 AVG button

8 HOLD button

+ -

A L A R M

E X I T

M E N U

1

S T O R E

2

L O G

3

R E C A L L

4

P R I N T

5

M A X

6

M I N

7

A V G

8

H O L D

9

S C A N

°

F

0

C L E A R

.

7

8

7

A V G

7 AVG button
Push to display the average temperature for each channel and the
number of readings that are being averaged. The display continues to
scan channels.
When a numerical entry is required and the NUM icon is on, push to
enter a ‘7’.

“AVG” appears and remains in the display.

Readings from when the instrument was last powered up or
cleared are averaged.

Opened channel readings are not averaged.

Use Left-Right arrow keys to stop scanning and to choose the
channel whose average is displayed. Push SCAN to return to
scanning channels while remaining in AVG.

Push AVG again to exit.

8 HOLD button
Push to hold the current record in display.
When a numerical entry is required, and the NUM icon is on, push to
enter an ‘8’.

“HOLD” appears and remains in the display.

Channel scanning, logging, printing continue normally but the
display remains.

Push HOLD or EXIT to exit.