Front panel displays and controls continued

How to adjust the log rate

NUM

LOG

How to adjust the log rate
The log rate is the time the instrument waits between logging temperature
scans in memory.
4,680 12 channel scans can be logged in memory, then LOG FULL is
displayed. When LOG FULL is displayed you can RECALL – scroll through
the logged readings, and you can PRINT – print the entire list of 4,680
logged entries. To clear (erase) memory: push CLEAR, then push LOG.
The default setting is 00:20 (20 seconds).
It takes 4 seconds to do one 12 channel scan so a log rate of 00:04 is the
maximum obtainable. The longest interval that can be set is 1 hour, 60:00.
Examples: At a setting of (00:20) the log is full in 26 hours. At the mini-
mum setting (00:04) the log is full in 5.2 hours. For 8 hours use a setting of
00:09. At the maximum setting (60:00) the log is full in 4,680 hours (195
days).
Manually pushing STORE adds one scan to the same memory as LOG.

(continued on next page)