Continued on next page) – Nova-Tech Digi–Sense 12-Channel Benchtop Scanning Thermometer User Manual

Page 52

Advertising
background image

51

Front panel displays and controls continued

How to do a field calibration continued

+ -

A L A R M

E X I T

MENU

1

S T O R E

2

L O G

3

R E C A L L

4

P R I N T

5

M A X

6

M I N

7

A V G

8

H O L D

0

C L E A R

9

S C A N

NUM

°

F

CAL

5

2a

2, 5

2b

2e

3

2d

2c

4

To do a field calibration continued:

1. Place the probe(s) at the desired temperature.
2. Push and release MENU until the display reads CALPT1, 2a.

The entry point 2b flashes.
CAL 2c is displayed. The channel number 2d is displayed.
NUM 2e appears and flashes indicating number keys 3 are active.

3. Push number keys to enter the temperature of the first calibration

point. (For example, 32.0 is entered by pushing 3 - 2 - 0.)

4. Use Left/Right keys to select next channel. Do all CALPT1's first.
5. Push MENU to switch to CALPT2.

CALPT 2 is skipped if no changes are made in CALPT1.

6. Place the probe(s) at the desired temperature.
7. Push and release number keys 7 to enter the temperature of the

second calibration point. (For example, 212.0 is entered by
pushing 2 - 1 - 2 - 0.)

+ -

A L A R M

E X I T

MENU

1

S T O R E

2

L O G

3

R E C A L L

4

P R I N T

5

M A X

6

M I N

7

A V G

8

H O L D

9

S C A N

NUM

°

F

CAL

5

0

CLEAR

7

(continued on next page)

BA 50-59

10/12/01, 11:07 AM

51

Advertising