Front panel displays and controls continued
How to adjust HI alarm setpoints

NUM

ALARM

°

F

HI

3

How to adjust HI alarm setpoints
The alarm mode must be ALARMS ON. If the alarm mode is ALARMS
OFF this setting is skipped in the MENU.
Only channels whose HI alarms are set to 1 (on) can have their
setpoints changed.

+ -

A L A R M

E X I T

MENU

1

S T O R E

2

L O G

3

R E C A L L

4

P R I N T

5

M A X

6

M I N

7

A V G

8

H O L D

9

S C A N

NUM

ALARM

°

F

HI

3

1a

1, 4b

4a

1d

1f

1c

3

1b

1e

2

0

CLEAR

To adjust a HI alarm setpoint:

1. Push MENU until SETPT 1a, the channel number 1b, the

setpoint value 1c, ALARM 1d, and HI 1e are displayed.
NUM 1f appears and flashes indicating number keys 3 are
active.

2. Push Left/Right arrow keys to select channel(s).
3. Push number keys to set the HI alarm temperature.

(For example, 156.8 is entered by pushing 1 - 5 - 6 - 8.)
Repeat steps 2 and 3 for all HI alarm set points you wish to
adjust.

4a. Push EXIT to save the setting and return to scanning, or

4b. Push MENU to switch to the next setup topic (adjust LO alarm

setpoints).