3 transmitter display and keypad, 4 password protection (default = 0) – Orion Instruments Eclipse 705 User Manual

Page 20

Advertising
background image

20

57-600 Eclipse

®

Guided Wave Radar Transmitter

2.6.3 Transmitter Display and Keypad

The ECLIPSE transmitter has an optional liquid crystal display
(LCD) capable of showing two lines of 8 characters each.
Transmitter measurements and configuration menu screens
are shown on the LCD.

The transmitter default display is the measurement screen.
It cycles every 5 seconds to display STATUS, LEVEL,
%OUTPUT, and LOOP information (LEVEL,
%OUTPUT, and STATUS for Fieldbus version). The
transmitter defaults to this display after 5 minutes if no
keystrokes are sensed.

The keypad has three arrows used to scroll through the dis-
plays and to calibrate the transmitter. The Up and Down
Arrow (

) keys and the Enter (

) key.

Function in

Function in

Arrows

Display Mode

Configuration Mode

Up and Down Moves forward and backward

Increases or decreases the

in the configuration program

value displayed or moves to

from one display to another.

another choice.

NOTE: Hold arrow key for
rapid scrolling.

Enter

Enters the configuration mode

Accepts a value and moves

(noted by an exclamation point to the next step of the
as the last character in the top

configuration program.

display line).

2.6.4 Password Protection (Default = 0)

The ECLIPSE transmitter is password protected to restrict
access to certain portions of the menu structure that affect
the operation of the system. When the proper password is
entered, an exclamation point (!) appears as the last charac-
ter of the first line of the display. The password can be
changed to any numerical value up to 255. The password is
required whenever configuration values are changed.

The default user password installed in the transmitter at the
factory is 0. The last step in the configuration menu pro-
vides the option to enter a new password. With a password
of 0, the transmitter is no longer password protected and
any value in the menu can be adjusted without entering a
confirming password, except diagnostic values.

NOTE: If the password is not known, the menu item New Password

displays an encrypted value representing the present pass-
word. Call the factory with this encrypted value to determine
the present password.