2.6.5.1 Measurement Type: Level Only (Loop Control = Level)

Display

Action

Comment

12

Senstvty

xxx

Enter value upward or
downward to sense liquid
surface

Allows fine gain adjustment for single rod probes (this parameter
is password protected for coaxial and twin rod probes).

13

LoopCtrl
(select)

Select variable to control
loop current

Select Level

14

Set 4mA

xxx.x 1u

Enter the PV value for the
4 mA point

A small transition zone (0–6") may exist at the top/bottom of
the probe. See Functional Specifications Probe, Section 3.6.1

15

Set 20mA
xxx.x lu

Enter the PV value for the
20 mA point

A small transition zone (0–6") may exist at the top/bottom of
the probe. Top 4" (100 mm) of 7xB Twin Rod Probe is inactive.
See Functional Specifications Probe, Section 3.6.1

16

Damping

xx s

Enter time constant of
desired damping

A Damping factor (0–10 seconds) may be added to smooth the
output due to turbulence

17

Fault

(select)

Select the loop current value
in presence of a fault

Select from 3.6 mA, 22 mA or HOLD

18

BlockDis

xx.x lu

Enter distance below refer-
ence point where level is not
sensed

Allows user to ignore level measurements near the top of the
probe

19

SZ Fault
(select)

Select loop current behavior
when level is sensed in safety
zone

Safety Zone is a user-defined area just below the Blocking
Distance
. Enable Fault if necessary to ensure safe, reliable high-
level readings in critical applications. Choices are None, 3.6 mA,
22 mA, Latch 3.6 or Latch 22. If Latch 3.6 or Latch 22 is
selected, the loop current will remain in alarm until it is manually
cleared with the SZ Alarm Reset below (#21)

20

SZ Height
(xx.x lu)

Enter distance below
BlockDis where SZ Fault will
be asserted

Enter a distance value that develops a safety zone just below the
Blocking Distance. Here the unit will report a Safety Zone Fault
(#19) if the level rises into this area.

21

SZ Alarm Reset

Press Enter to clear a
latched Safety Zone alarm

Clear a latched Safety Zone alarm

22

Threshld
(select)

Select the type of threshold

Unit default CFD. Only select Fixed in application with low
dielectric material over higher dielectric material and unit is read-
ing incorrect level. Example: Oil over water. (Adjustment of Trim
Level
may be necessary when threshold is changed)

23

Poll Adr

xx

Enter HART polling address
number (0-63)

Select a HART poll address (0–63). Enter 0 for a single
transmitter installation. Poll address does not affect loop current.

24

Loop Mode

Enable/Disable

Determines whether the loop is fixed at 4.0 mA or controlled by
the PV.

25

Trim Lvl

xx.x lu

Enter value to adjust Level
reading

-10.0 inches

≤ Lvl Trim ≤ +10.0 inches

(Requires superuser password)

26

Trim 4

xxxx

Fine tune the 4 mA point

Adjust setting to output exactly 4.0 mA on current meter

27

Trim 20

xxxx

Fine tune the 20 mA point

Adjust setting to output exactly 20.0 mA on current meter

28

Loop Tst

xx.x mA

Enter a mA Output value

Set mA output to any given value to perform loop test

29

LvlTicks

Xxxxx

Diagnostic Display

Time of flight from fiducial to level signal

30

New Pass

xxx

Enter new password (0-255)

Displays encrypted value of present password

31

Language
(select)

Select from English, Spanish,
French, German

Language choice for LCD display

32

Mdl705HT
Ver3.0a0

Transmitter display

Product identification Firmware version