4 measurement type: interface and volume – Orion Instruments Eclipse 705 User Manual

Page 32

Advertising
background image

32

57-600 Eclipse

®

Guided Wave Radar Transmitter

2.6.5.4 Measurement Type: Interface and Volume

Display

Action

Comment

46

Hist Rst

Press Enter and select yes
to clear history

Similar to SZ Alarm Reset

47

FidTicks

xxxx

Diagnostic Display

Time of flight from start of ramp to fiducial

48

Fid Sprd

49

Fid Type
(select)

Superuser Parameter

Select from positive or negative
(Selection only allowed for some probes, fixed for others)

50

Fid Gain

xxx

Superuser Parameter

51

Window

xxx

Factory Parameter

52

Conv Fct

xxxx

Factory Parameter

Calibration parameter

53

Scl Ofst

xxx

Factory Parameter

Calibration parameter

54

Neg Ampl

xxx

Superuser Parameter

55

Ifc Ampl

Superuser Parameter

56

Pos Ampl

xxx

Superuser Parameter

57

Signal

xxx

Diagnostic Display

Indication of level signal amplitude

58

Compsate
(select0

Superuser Parameter

Select from None, Manual, Auto

59

7xKCorr

xxx

Superuser Parameter

Distance in mm from fiducial to user reference point
(7xK probe characteristic)

60

ElecTemp

xxx C

Diagnostic Display

Present temperature in electronics compartment
(degrees Celsius)

61

Max Temp

xxx C

Superuser Password

Maximum electronics temperature recorded

62

Min Temp

xxx C

Superuser Password

Minimum electronics temperature recorded

63

SZ Hyst
xx.x lu

Superuser Password

Diagnostic factory setting