FM

705-5XXX-1XX

Intrinsically Safe

Class I, Div. 1; Groups A, B, C, & D

705-5XXX-2XX

Class II, Div. 1; Groups E, F, & G T4

Class III, Type 4X, IP66

Entity

705-5XXX-3XX

Explosion Proof

Œ

Class I, Div. 1; Groups B, C & D

705-5XXX-4XX

(with Intrinsically Safe probe)

Class II, Div. 1; Groups E, F, & G T4

Class III, Type 4X, IP66

705-5XXX-XXX

Non-Incendive

Class I, Div. 2; Groups A, B, C, & D

705-5XXX-XXX

Suitable for:



Class II, Div. 2; Groups F & G T4

Class III, Type 4X, IP66

CSA

705-5XXX-1XX

Intrinsically Safe

Class I, Div. 1; Groups A, B, C, & D

705-5XXX-2XX

Class II, Div. 1; Group E, F & G T4

Class III, Type 4X

Entity

705-5XXX-3XX

Explosion Proof

Œ

Class I, Div. 1; Groups B, C, & D

705-5XXX-4XX

(with Intrinsically Safe probe)

Class II, Div. 1; Group E, F & G T4

Class III, Type 4X

705-5XXX-XXX

Non-Incendive

Class I, Div. 2; Groups A, B, C, & D

705-5XXX-XXX

Suitable for:



Class II, Div. 2; Group E, F & G T4

Class III, Type 4X

IEC

705-5XXX-AXX

Intrinsically Safe

Ž

Zone 0 Ex ia IIC T4

705-5XXX-BXX

ATEX

705-5XXX-AXX

Intrinsically Safe

Ž

II

1G, EEx ia IIC T4

705-5XXX-BXX

705-5XXX-CXX

Flame Proof

II

1/2G, EEx d [ia] IIC T6

705-5XXX-DXX

705-51XX-EXX

Non-sparking

II

3(1)G, EEx nA [ia] IIC T4..T6

705-51XX-FXX

with probe II 1 G EEx ia IIC T6

705-52XX-EXX

II

3(1)G, EEx nA [nL] [ia] IIC T4..T6

705-52XX-FXX

with probe II 1 G EEx ia IIC T6

3.4

Agency Approvals

AGENCY

MODEL APPROVED

APPROVAL CATEGORY

APPROVAL CLASSES

48

57-600 Eclipse

®

Guided Wave Radar Transmitter

Œ

Factory Sealed: This product has been approved by Factory Mutual Research (FM), and Canadian Standards Association (CSA), as
a Factory Sealed device.



IMPORTANT:

Measured media inside vessel must be non-flammable only. If media inside vessel is flammable, then the

explosion proof version (which contains an internal barrier making the probe Intrinsically Safe) is required.

Ž

Special conditions for safe use
Because the enclosure of the Guided Wave Radar Level Transmitter ECLIPSE Model 705-5 _ _ _ - _1_ and/or Probe ECLIPSE Model
7_ _ -_ _ _ _ - _ _ _ is made of aluminum, if it is mounted in an area where the use of category 1 G (Zone 0) apparatus is required, it
must be installed such, that, even in the event of rare incidents, ignition sources due to impact and friction sparks are excluded.

For applications in explosive atmospheres caused by gases, vapors or mists and where category 1G (Zone 0) apparatus is required,
electrostatic charges on the non-metallic parts of the Probe ECLIPSE Model 7x5- _ _ _ _ - _ _ _ , Model 7x7- _ _ _ _ - _ _ _ and
Model 7_ F- _ _ _ _ - _ _ _ shall be avoided.

These units are in conformity of:
1. The EMC Directive: 2004/108/EC. The units have

been tested to EN 61326.

2. Directive 94/9/EC for equipment or protective system

for use in potentially explosive atmospheres.

0344

Note: Single and twin rod probes must be used in metallic

vessel or stillwell to maintain CE compliance.

3.4.1

Agency Specifications – Explosion Proof Installation

Factory Sealed: This product has been approved by Factory Mutual Research (FM), and, Canadian Standards
Association (CSA), as a Factory Sealed device.

NOTE: Factory Sealed: No Explosion Proof conduit fitting (EY seal) is required within 18" of the transmitter. However, an
Explosion Proof conduit fitting (EY seal) is required between the hazardous and safe areas.

Caution: Grounding (+) will cause faulty operation, but will not cause permanent damage.