3.3.2 Status Messages

Display Message

Action

Comment

Lo Temp

Present temperature in
electronics compartment is
below -40° C

1) Transmitter may need to be moved to ensure ambient temperature

is within specification

2) Change to remote mount transmitter

HiVolAlm

Level more than 5% above
highest point in strapping table

Verify strapping table is entered correctly. None. Signal amplitude is
lower than desired.

Sys Warn

Unexpected but non-fatal
software event

Consult factory

TrimReqd

Factory set Loop values are
defaults, loop output may be
inaccurate

Consult factory

Cal Reqd

Factory set default calibration
parameters are in use, level
reading may be inaccurate

Consult factory

SlopeErr

Ramp circuit generating
improper voltage

Consult factory

LoopFail

Loop current differs from
expected value

Consult factory

No Ramp

No End-of-Ramp signal detected Consult factory

DfltParm

Internal non-volatile parameters
have been defaulted

Consult factory

LVL < Probe

Length

Apparent position of the upper
level pulse is beyond the end of
probe.

1) Check entered probe length
2) Change threshold to fixed

EE Fail

EEPROM error allowing watch-
dog timer to expire

Consult factory

CPU Fail

A-D converter time out allowing
watchdog timer to expire

Consult factory

SfwrFail

A fatal software error allowing
watchdog timer to expire

Consult factory

PACTware

PC Program

The ECLIPSE Model 705 offers the ability to do Trending and Echo Curve analysis using a PACTware DTM.
This is a powerful troubleshooting tool that can aid in the resolution of some of the Error Messages shown above.

Refer to Bulletins 59-101 and 59-601 for more information.