Automation Command and Query Reference Manual - Control Reference

CombinedChannels

Enum

Select MS-500 operation mode. In "2Combine", sampling rate is up to 2 GS/s and you can use up to
18 digital input lines. In "NoCombine", sampling rate is up to 1 GS/s and you can use up to 36 digital
input lines.

Description

Values

2Combine

NoCombine

LineNames

String

Any number of characters

Range

List of Digital Line Names

Description

MSxxHysteresis0

Double

From 0.5 to 0.5 step 0.02

Range

The minimum high voltage level is user definable by the hysteresis control up to 1.4 V above the
threshold. The maximu low voltage level is user definable by the hysteresis control up to 1.4 V below
the threshold. The minimum hysteresis is 100 mV.

Description

MSxxHysteresis1

Double

From 0.5 to 0.5 step 0.02

Range

MSxxHysteresis2

Double

From 0.5 to 0.5 step 0.02

Range

MSxxHysteresis3

Double

From 0.5 to 0.5 step 0.02

Range

MSxxThreshold3

Double

1-120

918501 RevA