Automation Command and Query Reference Manual - Control Reference

HorPos

Double

From -0.5 to 0.5 step (8 digits)

Range

Sets/Queries the horizontal position of center of the grid on the zoomed trace Fx. The unit of
measurement is the screen width, that is, 0.3 means a shift of three of the ten divisions. A positive
value moves the trace to the left.

Description

HorZoom

Double

From 0.1 to 1e+006 step (8 digits)

Range

Sets/Queries the horizontal magnification of the trace Fx. The magnification will be in a 1 2 5 10
sequence unless variable horizontal magnification has been set.

Description

NumSelectedSegments

DoubleLockstep

From 1 to 1 step 1, locked to 1 2 5, fine grain allowed=true, on=false

Range

VariableHorZoom

Bool

Sets/Queries the ability to zoom horizontally by a continuously variable factor. Note that if a horizontal
zoom of 0.9 is set, while variable zoom is off, the horizontal zoom will be set to 1.0. If the variable
zoom is then enabled, the factor of 0.9 will have been remembered, and it will be used. Note that the
previous value will not be remembered during a power-cycle.

Description

VariableVerZoom

Bool

Sets/Queries the ability to zoom vertically by a continuously variable factor. Note that if a vertical
zoom of 0.9 is set, while variable zoom is off, the vertical zoom will be set to 1.0. If the variable zoom
is then enabled, the factor of 0.9 will have been remembered, and it will be used. Note that the
previous value will not be remembered during a power-cycle.

Description

VerPos

Double

From -1.5 to 1.5 step (8 digits)

Range

Sets/Queries the vertical position of center of the grid on the zoomed trace Fx. The unit of
measurement is the screen height, that is, 0.375 means a shift of three of the eight divisions. A
positive value moves the trace downwards.

Description

VerZoom

Double

From 0.1 to 100 step (8 digits)

Range

Sets/Queries the vertical magnification of the trace Fx. The magnification will be in a 1 2 5 10
sequence unless VariableVerZoom has been set to True, in which case it will be continuously variable.

Description

XY

app.Math.XY

This set of variables controls the display of data in X vs. Y mode. Only Valid when the instrument is in XY, XYSingle,
or XYDual display modes.

1-150

918501 RevA