Automation Command and Query Reference Manual - Control Reference

RiseTime3

Double

From 0 to 2e-009 step 1e-012

Range

When probes are deskewed, this control is set to the measured risetime of Channel 3.

Description

RiseTime4

Double

From 0 to 2e-009 step 1e-012

Range

When probes are deskewed, this control is set to the measured risetime of Channel 4.

Description

Skew1

Double

From -4.5e-007 to 4.5e-007 step 1e-012

Range

Defines the skew value for Channel 1. Units are in seconds. A positive value has the effect of moving
the trace to the right.

Description

Skew2

Double

From -4.5e-007 to 4.5e-007 step 1e-012

Range

Defines the skew value for Channel 2. Units are in seconds. A positive value has the effect of moving
the trace to the right.

Description

Skew3

Double

From -4.5e-007 to 4.5e-007 step 1e-012

Range

Defines the skew value for Channel 3. Units are in seconds. A positive value has the effect of moving
the trace to the right.

Description

Skew4

Double

From -4.5e-007 to 4.5e-007 step 1e-012

Range

Defines the skew value for Channel 4. Units are in seconds. A positive value has the effect of moving
the trace to the right.

Description

SkewCal1

Double

From -4.5e-007 to 4.5e-007 step 1e-012

Range

SkewCal2

Double

From -4.5e-007 to 4.5e-007 step 1e-012

Range

SkewCal3

Double

From -4.5e-007 to 4.5e-007 step 1e-012

Range

1-197