Automation Command and Query Reference Manual - Control Reference

FindLevel

Action

When this action control is used, SDA2 will set the PercentLevel to 50% and will set the AbsLevel to
halfway between the base and top of the input signal.

Description

InputSUT

Enum

This control specifies whether the input Signal Under Test is single ended (from one source, such as
C2) or differential (from a pair of sources, such as C2 and C3).

Description

Values

Diff

Differential signal, a "plus" and "minus" side, from two sources; SDA2 will calculate the difference.

Single

Single ended signal, in one source

LevelType

Enum

This control sets whether PercentLevel or AbsLevel is used to define the threshold for detecting
transitions in the input signal.

Description

Values

Absolute

Percent

PercentLevel

Double

From 0 to 100 step 1

Range

When LevelType is set to Percent, this control specifies the threshold level. Its value is percent from
base (0%) to top (100%).

Description

1-374