VCBA VERSION 10.xx SOFTWARE MANUAL

REV 2

45

6. Click on the “OK” button. The override value will be inserted in the tabulated test results

as shown:

The above method works well for simple tests such as an OPEN or CLOSE test. For more
complex situations such as a CLOSE-OPEN test with an insertion resistor, you can manually
input the override values. To manually input the override values:

1. Retrieve the test record that you would like to view (See sections 4.2 and 4.3). The

tabulated test results will be displayed.

2. Press the [F8] key in either the tabulated results view or the graphic results view.

3. The “Contact Override” window will be displayed as in the previous example above.

Select the contact number from the “Contact:” dropdown list.

4. Click on the “Enable” checkbox next to each value you would like to override for the

selected contact, and then enter the override value in the input area to the left of the
checkbox.

5. Click on the “OK” button and all enabled override values will be inserted in the

tabulated test results.