2 zero point adjustment, Zero point adjustment -2, Important – Yokogawa EJX210B User Manual

Page 30

Advertising
background image

<7. Operation>

7-2

IM 01C27C01-01EN

• Measurement range (measurement lower/

upper limit, unit).

7.2 Zero Point Adjustment

After completing preparations for operating the

transmitter, adjust the zero point. There are two

zero point adjusting ways.

IMPORTANT

Do not turn off the power to the transmitter

immediately after performing a zero point

adjustment. Powering off within 30 seconds of

performing this procedure will return the zero

point to its previous setting.

(1) When you can obtain Low Range Value

from actual measured value of 0% (0 kPa,

atmospheric pressure);

F0703.ai

Zero-adjustment

screw cover

Figure 7.3

External Zero Adjustment Screw

■ Using the Zero-adjustment screw
Before adjusting zero point, make sure followings.

• The External zero trim parameter

(External Zero Trim) is “Trim on”. For details,

refer to section 8 “Setting Parameters”.

• Use a slotted screwdriver to turn the zero-

adjustment screw. Turn the screw clockwise

to increase the output or counterclockwise to

decrease the output.

The zero point adjustment can be made with

a resolution of 0.01% of the setting range. The

degree of zero adjustments varies with the

screw turning speed; turn the screw slowly

to make a fine adjustment, quickly to make a

rough adjustment.

■ Using the Device Configuration Tool
Refer to subsection 8.3.11 “Zero Point Adjustment

and Span Adjustment”.

(2) When you cannot obtain Low Range Value

from actual measured value of 0%;

When adjusting the transmitter zero point, the liquid

level in a tank can not be set to the low limit (0%) of

the measuring range; use a digital manometer or a

glass gauge to match the transmitter output signal

with the actual measured value.
Adjust the transmitter output to the actual measured

value obtained by a digital manometer or a glass

gauge.
[Example]
The measuring range of 0m to 2m; the actual

measured value of 0.8m.

0.8

2

Actual measured value=

x100=40.0%

■ Using the Zero-adjustment screw
Turn the zero adjustment screw to match the output

signal to the actual measured value.

■ Using the Device Configuration Tool
Refer to subsection 8.3.11 “Zero Point Adjustment

and Span Adjustment”.

Advertising