Yokogawa EJX210B User Manual

Page 89

Advertising
background image

<11. General Specifications>

11-2

IM 01C27C01-01EN

Zero Adjustment Limits:

Zero can be fully elevated or suppressed, within

the lower and upper range limits of the capsule.

External Zero Adjustment:

External zero is continuously adjustable with

0.01% incremental resolution of span. Re-range

can be done locally using the digital indicator

with range-setting switch.

Integral Indicator (LCD display):

5-digit numerical display, 6-digit unit display and

bar graph.

The indicator is configurable to display one

or up to three of the following variables

periodically.; Differential pressure, static

pressure, temperature.

EMC Conformity Standards

EN61326-1 Class A, Table 2 (For use in

industrial locations), EN61326-2-3

R&TTE Conformity Standards

ETSI EN 300 328, ETSI EN 301 489-1,

ETSI EN 301 489-17, EN61010-1,

EN61010-2-030, EN62311

• Indoor/Outdoor use

Regulation Conformity of the Wireless Module

• FCC Approval

• IC Approval

Safety Requirement Standards

EN61010-1, EN61010-2-030

• Altitude of installation site:

Max. 2,000 m above sea level

• Installation category: I

(Anticipated transient overvoltage 330 V)

• Pollution degree: 2

• Indoor/Outdoor use

Process connections:

High pressure side:

Flange connected

See the following table.

Table 11.2

Flange size and rating

Process

connection style

Size

Flange

Flush type

3-inch

2-inch

1 1/2-inch*

JIS 10K, 20K

ANSI Class 150, 300

JPI Class 150, 300

DIN PN 10/16, 25/40

Extended type

4-inch

3-inch

JIS 10K, 20K

ANSI Class 150, 300

JPI Class 150, 300

DIN PN 10/16, 25/40

*:

Flushing connection rings are always attached.

Low pressure side:

Threaded

See “Model and Suffix Codes.”

Process connection of cover flange: IEC61518

Gasket Contact Surface:

See the following table.

Table 11.3

Gasket contact surface

Flange

JIS/JPI/DIN

ANSI

Wetted parts material code

SW,

SE,

WW,

WE

HW,

TW

SW,

SE,

WW,

WE

HW,

TW

Gasket cntact

Surface

Serration

*1

Flat (No serration)

●: Applicable, —: Not applicable
*1:

ANSI B16.5

Table 11.1

Process temperature, Ambient temperature, and Working pressure

Code

Process temperature

*1*2

Ambient temperature

*3

Working pressure

Silicone oil

A

-10 to 250°C

*4

(14 to 482°F)

-10 to 85°C

(14 to 185°F)

2.7 kPa abs

(0.38 psi abs) to

flange rating pressure

*1:

See figure 11.1 ‘Working Pressure and Process Temperature.’

*2:

Indicates high pressure side value. The process temperature limit for low pressure side is –40 to 120°C (–40 to 248°F).

*3:

This ambient temperature is the transmitter ambient temperature.

*4:

In case of wetted parts material code TW (Tantalum), process temperature limit is up to 200°C (392°F).

Advertising