Diagnostic Results Summary

Diagnostic Status

Contents

Alert

Type

NAMUR

NE107

Category

Diagnostic Status Detail

Description

Faults in electronics

78

F

AMP_T_SENSOR_FAIL*

Amplifier temperature sensor failure

AMP_EEPROM_FAIL

Amplifier EEPROM failure

AMP_EEP_IRREGULAR

AMP EEPROM version not correct

G_A_COMM_FAIL

G/A failure

FC_DELTA_T_FAIL

C-side delta T circuit failure

FR_DELTA_T_FAIL

R-side delta T circuit failure

WL_AD_FAIL

Battery voltage not detected (AMP

failure)

Faults in sensor or actuator

element

77

F

FC_SENSOR_FAIL

C sensor frequency failure

FR_SENSOR_FAIL

R sensor frequency failure

CAP_T_SENSOR_FAIL

Capsule temperature sensor failure

CAP_EEPROM_FAIL

Capsule EEPROM failure

CAP_EEP_IRREGULAR

CAP EEPROM version not correct

FC_UNOSC_FAIL

C sensor oscillation stop failure

FR_UNOSC_FAIL

R sensor oscillation stop failure

Installation, calibration

problem

76

C

DP_TRIM_SPAN_OUTSIDE "Pressure span adjustment variable

outside of range"

DP_TRIM_ZERO_OUTSIDE "Pressure zero adjustment variable

outside of range"

SP_TRIM_SPAN_OUTSIDE "Static pressure span adjustment

variable outside of range"

SP_TRIM_ZERO_OUTSIDE "Static pressure zero adjustment

variable outside of range"

LCD_OUTSIDE_LIMIT

LCD display outside of limits

Out of service

75

O

AI1_OUT_OF_SERVICE

AI1 O/S mode

AI2_OUT_OF_SERVICE

AI2 O/S mode

AI3_OUT_OF_SERVICE

AI3 O/S mode

Outside sensor limits

74

C

DP_OUTSIDE_LIMIT

Pressure outside of range

SP_OUTSIDE_LIMIT

Static pressure outside of range

CAPT_OUTSIDE_LIMIT

Capsule temperature outside of range

AMPT_OUTSIDE_LIMIT

Amplifier temperature outside of range

Environmental conditions

out of device specification.

73

O

DP_OUTSIDE_RANGE

Pressure setting outside of range

SP_OUTSIDE_RANGE

Static pressure setting outside of

range

Power is critical low:

maintenance need short-

term.

71

M

WL_DEEPSLP_ALM *

Deep sleep due to low battery

CRITICAL_LOWBAT **

Deep sleep due to low battery

Power is low:

maintenance need mid-term 70

M

WL_LOWBAT_ALM

Low battery

Simulation is active

68

C

AI1_SIMULATION_ACTIVE

AI1 Simulation mode

AI2_SIMULATION_ACTIVE

AI2 Simulation mode

AI3_SIMULATION_ACTIVE

AI3 Simulation mode

Not applicable for the diagnostic regarding AI2 object and static pressure measurement.

*:

Applicable for Integral antenna type (Amplifier housing code: 7).

**:

Applicable for Detachable antenna type (Amplifier housing code: 8 or 9).

Advertising