4 disassembly and reassembly, 1 replacing the integral indicator, Disassembly and reassembly -3 9.4.1 – Yokogawa EJX210B User Manual

Page 65: Replacing the integral indicator -3, Caution, Important

Advertising
background image

<9. Maintenance>

9-3

IM 01C27C01-01EN

Table 9.2

Tools for Disassembly and Reassembly

Tool

Quantity

Remarks

Phillips

screwdriver

1

JIS B4633, No. 2

Slotted

screwdriver

1

Allen wrenches

3

JIS B4648

One each, nominal 3,

4 and 2.5 mm Allen

wrenches

Wrench

1

Width across flats, 17 mm

Torque wrench

1

Adjustable

wrench

1

Socket wrench

1

Width across flats, 16 mm

Socket driver

1

Width across flats, 5.5 mm

Tweezers

1

9.4.1 Replacing the Integral Indicator

This subsection describes the procedure for

replacing an integral indicator. (See figure 9.2)

■ Removing the Integral Indicator
1) Remove the cover.

2) While supporting the integral indicator with one

hand, loosen its two mounting screws.

3) Dismount the integral indicator from the RF

assembly.

When doing this, carefully pull the integral

indicator straight forward so as not to damage

the connector pins between it and the RF

assembly.

■ Attaching the Integral Indicator
1) Align both the integral indicator and RF

assembly connectors and engage them.

2) Insert and tighten the two mounting screws.

3) Replace the cover.

9.4 Disassembly and

Reassembly

This section describes procedures for disassembly

and reassembly for maintenance and component

replacement.

CAUTION

Precautions for the intrinsic safe type

transmitters
Intrinsic safe type transmitters must be, as

a rule, removed to a non-hazardous area

for maintenance and be disassembled and

reassembled to the original state. Check and

confirm the insulation when it is reassembled to

the original state.

Check and confirm the insulation when it is

reassembled to the original state.

Refer to section 2.7 “Insulation Resistance and

Dielectric Strength Test” for details of Resistance

Test.

IMPORTANT

• Perform the provisioning when replacing the

RF assembly. Refer to 7.4 Connecting to the

Field Wireless Network for details.

• Replace the batteries and perform the

parameter settings when replacing the CPU

assembly. Refer to 9.4.6 Replacing the

Batteries.

CAUTION

Always remove Battery pack and shut off

pressures before disassembly and assembly.

Use proper tools for all operations.

Table 9.2 shows the tools required.

Advertising