Gonogo:script, Template:mode, Template:maximum – GW Instek GDS-3000 Series Programming Manual User Manual
Page 100

GDS-3000 Programming Manual
100
:GONogo:SCRipt
Set
Description
Activates/Deactivates the Go/NoGo APP or
queries its state.
Syntax
:GONogo:SCRipt {OFF | ON | ?}
Parameter/
Return Parameter
ON
Turns Go/NoGo APP on.
OFF
Turns the Go/NoGo APP off.
Example
:GONogo:SCRipt?
>ON
The Go/NoGo script is on.
:TEMPlate:MODe
Set
Query
Description
Sets or returns the Go/NoGo template mode.
Syntax
:TEMPlate:MODe{MAXimum|MINimum|AUTO|?}
Parameter/
Return Parameter
MAXimum
Maximum template
MINimum
Minimum template
AUTO
Auto template
Example
:TEMPlate:MODe
AUTO
Sets the template mode to AUTO.
:TEMPlate:MAXimum
Set
Query
Description
Defines or queries which waveform memory
(REF1 or W1~W20) is set to the maximum
template.
Syntax
:TEMPlate:MAXimum{REF1|W1~W20|?}
Parameter/
Return Parameter
REF1
Reference one
W1~W20 Waveform memory 1 to 20