Acquire<x>:state – GW Instek GDS-3000 Series Programming Manual User Manual
Page 35

COMMAND DETAILS
35
Example
:ACQuire1:MEMory?
Memory Length,25000;IntpDistance,0;Trigger
Address,12499;Trigger Level,1.00V;Sou
rce,CH1;Vertical Units,V;Vertical Scale,5.000e-
01;Probe,1.000e+00;Vertical Posit
ion,-1.460e+00;Horizontal Units,S;Horizontal
Scale,5.000E-04;Horizontal Position
,0.000E+00;Horizontal Mode,Main;SincET Mode,Real
Time;Sampling Period,2.000e-07;
Horizontal Old Scale,5.000E-04;Horizontal Old
Position,0.000E+00;Firmware,V1.08;
Time,07-Feb-11 15:35:17;Waveform Data;#550000<50000
bytes binary data><LF>
:ACQuire<x>:STATe?
Query
Description
Returns the status of waveform data.
Syntax
:ACQuire<x>:STATe?
Parameter
<x>
Channel
1/2/3/4
Channel 1/2/3/4
Return parameter 0
Raw data is not ready
1
Raw data is ready
Example
:ACQuire1:STATe?
0
Returns 0. The channel 1’s raw data is not ready.
Note: If the oscilloscope changes the acquisition status
from STOP to RUN, the status will be reset as zero.