Gonogo:execute, Gonogo:function, Gonogo:ngcount – GW Instek GDS-3000 Series Programming Manual User Manual
Page 98: Gonogo:ngdefine

GDS-3000 Programming Manual
98
:GONogo:EXECute
Set
Query
Description
Enables or disables the Go/NoGo function or
queries its state.
Syntax
:GONogo:EXECute {OFF|ON|?}
Parameter/
OFF
Disabled
Return Parameter ON
Enabled
Example
:GONogo:EXECute OFF
Turns Go/NoGo off.
:GONogo:FUNCtion
Set
Description
Initializes the Go/NoGo APP. This must be run
after the Go/NoGo APP has been started.
Syntax
:GONogo:FUNCtion
:GONogo:NGCount
Query
Description
Returns the Go/NoGo counter.
Syntax
:GONogo:NGCount {?}
Return parameter Returns a string in the following format “number of
violations,total tests”
Example
:GONogo:NGCount?
> 3,25
Indicates that 3 violations occurred over 25 tests.
:GONogo:NGDefine
Set
Query
Description
Sets the Go/NoGo “When” conditions.
Syntax
:GONogo:NGDefine {EXITs|ENTers|?}