Trigger:llevel, Trigger:edge:slop, Trigger:delay:type – GW Instek GDS-3000 Series Programming Manual User Manual
Page 84

GDS-3000 Programming Manual
84
Syntax
:TRIGger:HLEVel {
<NRf> | ?}
Parameter
<NRf>
High level value
Return parameter
Returns the trigger high level as <NR3>.
Example
:TRIGger:HLEVel 3.30E-1
Sets the trigger high level to time to 330mV/mA.
:TRIGger:LLEVel
Set
Query
Description
Sets or queries the low trigger level (applicable for
Rise and Fall/Pulse Runt trigger).
Syntax
:TRIGger:LLEVel {
<NRf> | ?}
Parameter
<NRf>
Low level value
Return parameter
Returns the trigger low level as <NR3>.
Example
:TRIGger:LLEVel -3.30E-3
Sets the trigger low level to time to -330mV/mA.
:TRIGger:EDGe:SLOP
Set
Query
Description
Sets or queries the trigger slope.
Syntax
:TRIGger:EDGe:SLOP {RISe | FALL | ?}
Parameter
RISe
Rising slope
FALL
Falling slope
Return parameter
Returns the trigger slope.
Example
:TRIGger:EDGe:SLOP FALL
Sets the trigger slope to falling.
:TRIGger:DELay:TYPe
Set
Query
Description
Sets or queries trigger delay type.
Syntax
:TRIGger:DELay:TYPE {TIMe | EVENt | ?}