Channel<x>:bwlimit, Channel<x>:coupling, Channel<x>:deskew – GW Instek GDS-3000 Series Programming Manual User Manual
Page 37

COMMAND DETAILS
37
:CHANnel<x>:BWLimit
Set
Query
Description
Selects or returns the bandwidth limit on/off.
Syntax
:CHANnel<x>:BWLimit {FULL | <NR3> | ?}
Parameter
<x>
Channel
<NR3>
Limit
1/2/3/4 CH1/2/3/4
20E+6
20MHz
FULL
Full bandwidth 100E+6
100MHz
200E+6
200MHz
350E+6
350MHz
Return Parameter <NR3>
Returns the bandwidth.
Full
Full bandwidth for the oscilloscope.
Example
:CHANnel1:BWLimit 2.000E+07
Sets the channel 1 bandwidth 20MHz
:CHANnel<x>:COUPling
Set
Query
Description
Selects or returns the coupling mode.
Syntax
CHANnel<x>:COUPling {AC | DC | GND | ?}
Parameter
<x>
Channel
Coupling mode
1/2/3/4 CH1/2/3/4
AC
AC coupling
DC
DC coupling
GND
Ground coupling
Return parameter Returns the coupling mode.
Example
:CHANnel1:COUPling DC
Sets the coupling to DC for Channel 1.
:CHANnel<x>:DESKew
Set
Query
Description
Sets the deskew time in seconds.