Channel<x>:bwlimit, Channel<x>:coupling, Channel<x>:deskew – GW Instek GDS-3000 Series Programming Manual User Manual

Page 37

Advertising
background image

COMMAND DETAILS

37

:CHANnel<x>:BWLimit

Set

Query

Description

Selects or returns the bandwidth limit on/off.

Syntax

:CHANnel<x>:BWLimit {FULL | <NR3> | ?}

Parameter

<x>

Channel

<NR3>

Limit

1/2/3/4 CH1/2/3/4

20E+6

20MHz

FULL

Full bandwidth 100E+6

100MHz

200E+6

200MHz

350E+6

350MHz

Return Parameter <NR3>

Returns the bandwidth.

Full

Full bandwidth for the oscilloscope.

Example

:CHANnel1:BWLimit 2.000E+07
Sets the channel 1 bandwidth 20MHz

:CHANnel<x>:COUPling

Set

Query

Description

Selects or returns the coupling mode.

Syntax

CHANnel<x>:COUPling {AC | DC | GND | ?}

Parameter

<x>

Channel

Coupling mode

1/2/3/4 CH1/2/3/4

AC

AC coupling

DC

DC coupling

GND

Ground coupling

Return parameter Returns the coupling mode.

Example

:CHANnel1:COUPling DC
Sets the coupling to DC for Channel 1.

:CHANnel<x>:DESKew

Set

Query

Description

Sets the deskew time in seconds.

Advertising