Measure:gating, Measure:source<x, Measure:fall – GW Instek GDS-3000 Series Programming Manual User Manual
Page 59

COMMAND DETAILS
59
:MEASure:GATing
Set
Query
Description
Sets or queries the measurement gating.
Syntax
:MEASure:GATing {OFF|SCREen|CURSor|?}
Parameter
OFF
Full record CURSor
Gating
between
cursors
SCREen
Gating set
to screen
width
Return parameter
Returns the gating. (OFF, SCREEN, CURSOR)
Example
:MEASure:GATing OFF
Turns gating off (full record).
:MEASure:SOURce<x>
Set
Query
Description
Sets or queries the measurement source for source1
or source2.
Syntax
:MEASure:SOURce<x> {CH1|CH2|CH3|CH4?}
Parameter
<x>
Source1 or
source2
CH1~CH4
Channel 1 to
4
1,2
Return parameter
Returns the source (CH1, CH2, CH3,CH4)
Example
:MEASure:SOURce1 CH1
Sets source1 to channel 1.
:MEASure:FALL
Query
Description
Returns the fall time measurement result.
Syntax
:MEASure:FALL{?}
Return parameter <NR3>
Chan Off
Indicates the source channel is not
activated.