PEL-3000 User Manual
102
Memo
12 characters
A user-created note for the
currently selected sequence.
Mode
CC, CR, CV, CP
Operating mode for the
sequence. +CV mode is
supported.
Range
ILVL
Low I range, low V range
IMVL
Middle I range, low V range
IHVL
High I range, low V range
ILVH
Low I range, high V range
IMVH
Middle I range, high V range
IHVH
High I range, high V range
Loop
Infinite,
01 ~ 9999
Sets the amount of times to
loop the selected sequence.
Last Load
OFF, ON
Set the load condition after
the end of the sequence.
Last
Value
The setting value of the load
for when Last Load = ON.
Chain
Off, S01~S10
Sets the next sequence in the
chain, when not set to off.
Data Edit
Overview
Each step in a normal sequence contains the
following setting parameters:
Setting
Setting Range
Description
Step
0001 ~ 1000
Selects/displays the current
step in the sequence.
The number of available
steps is dependent on the
number of steps added using
the Insert Point[F1] functions.
Value
The current, voltage, power
or resistance setting for the
selected operating mode.
Load
ON, OFF
Turns the load on or off for
the selected step.