Parallel capacity, pel-3211, Connection – GW Instek PEL-3000 Series User Manual User Manual
Page 155

EXTERNAL CONTROL
153
Parallel Capacity, PEL-3211
Model No. of
Units
V
I
Total Sink Current
PEL-3111 +
PEL-3211
Total Power
PEL-3111 +
PEL-3211
PEL-3111:
Master
x 1 150V 210A
N/A
N/A
PEL-3211:
Slave
Boosters
x 1 150V 420A
630A
3150W
x 2 150V 840A
1050A
5250W
x 3 150V 1260A
1470A
7350W
x 4 150V 1680A
1890A
9450W
Note: The PEL-3211 booster packs do not have a control panel. They can
only be used as slaves with a single PEL-3111 in parallel.
Connection
Description
The J1 and J2 connectors are used for control
during parallel operation. Up to 5 units can be
used in parallel.
Note
Only the rear panel terminals can be used for
parallel operation, the front panel terminals have a
lower current rating and thus should not be used
for parallel operation.