Load Off (Mode) and Load Off (Range)
Description
By default the load will automatically turn off
when the either the operating mode (CC, CV,
CR, CP) or the range (I range, V range) is
changed.
To allow the load to stay on when the operating
mode is changed, set the Load Off (Mode) setting
to OFF.
To allow the load to stay on when the current
or voltage range is changed, set the Load Off
(Range) setting to OFF.
By default, these settings are set to ON.
Operation
1. Press
Shift
+
Utility
Help
> Load[F2].
2. Select Load Off (Mode) setting.
When set to OFF, the load will stay on when the
operating mode is changed.
Load Off (Mode):
OFF, ON
3. Select Load Off (Range) setting.
When set to OFF, the load will stay on when the
range is changed.
Load Off (Range):
OFF, ON