GW Instek PEL-3000 Series User Manual User Manual
Page 56

PEL-3000 User Manual
54
6. Set the resistance or conductance level
parameters using the scroll wheel and number
pad.
For Static mode, set CR A Value and/or CR B
Value.
For Dynamic mode, set Level1 and Level2.
The maximum and minimum conductance/
resistance levels depend on the selected current
range.
7. To add CV mode to CR mode (CR+CV), see
page 58.
8. Set the remaining basic configuration settings
such as the slew rate, and switching mode
settings. See page 63 for details.
Display
Mode
CR
I Range
H 35A
V Range
L 15V
Function
Static
Configure
0.000
V
0.000
A
0.00
w
SlewRate
250.000
CR B Value
0.04082
CR A Value
0.04082
LOAD
RS232
A Value
Fine
03/Sep/2012
mA/uS
Ω
Ω
Mode
Current range
Voltage range
Active setting
Conductance/
Resistances
Conductance/
Resistance settings
Note
Basic CR mode configuration is complete. See
page 63 for more configuration options.
The current range and voltage range applies to all
the operating modes.