CP Mode

Proportional Control:
Input power = rated power × (external
resistance/10).

Inverse Control:
Input power = rated power × (1 - external
resistance/10).

Input

Power

External

Resistance

Rated

Power

0

10k

Proportional control

Inverse control

Note

The inverse configuration is recommended for
safety reasons. In the event that any of the cables
become accidentaly disconnected, the
current/voltage/power input will drop to the
minimum. Under similar circumstances using
proportional control, an unexpectedly high input
would result.

Operation

1. Turn the power off from the PEL-3000 and from

the load.

2. Connect the external resistance across pins 1

and 3 of the J1 connector.

3. Turn the power on the PEL-3000.

4. Set the operating mode and range.

See page 51 for CC mode.

See page 53 for CR mode.

See page 55 for CV mode.

See page 56 for CP mode.