GW Instek PEL-3000 Series User Manual User Manual
Page 108

PEL-3000 User Manual
106
Data Edit Configuration
Data Edit Display
Previous
Menu
LOAD
RS232
03/Sep/2012
Data Edit for Normal Sequences
RAMP:
LOAD:
OFF
OFF
000 H:
PAUSE:
TRIG OUT
OFF
OFF
001ms
0001
Insert
Point
Save
Step:
0001
Time:
Value:
0.00000
A
/
00 M: 00 S.
Delete
Point
Active step number
Total number of steps
Step settings
Operation
1. Press
FUNC
> Sequence[F2] > Normal
Sequence[F1].
2. Select Seq.No. and select the sequence you wish
to edit.
Start:
S01 ~ S10
3. Press Edit Sequence [F2] to enter the Data Edit
configuration menu.
Note that when there no steps in the current
sequence the Data Edit for Normal Sequence
settings are blank.
4. Press Insert Point[F1] to add a step to the
sequence after the current step.
Every time Insert Point is pressed the Step
parameter is incremented.
The inserted point becomes the current step.