Cc mode – GW Instek PEL-3000 Series User Manual User Manual
Page 198

PEL-3000 User Manual
196
CC Mode
Model
PEL-3021
PEL-3041
PEL-3111
Operating Range
H Range
0A~35A
0A~70A
0A~210A
M Range
0A~3.5A
0A~7A
0A~21A
L Range
0A~0.35A
0A~0.7A
0A~2.1A
Setting Range
H Range
0A~35.7A
0A~71.4A
0A~214.2A
M Range
0A~3.57A
0A~7.14A
0A~21.42A
L Range
0A~0.357A
0A~0.714A
0A~2.142A
Default Setting
H Range
0A
0A
0A
M Range
0A
0A
0A
L Range
0A
0A
0A
Resolution
H Range
1mA
2mA
10mA
M Range
0.1mA
0.2mA
1mA
L Range
0.01mA
0.02mA
0.1mA
Accuracy of Setting
H, M Range ±(0.2 % of set + 0.1 % of f.s
*1
) + Vin
*2
/500 kΩ
L Range
±(0.2 % of set + 0.1 % of f.s) + Vin
*2
/500 kΩ
Parallel
Operation
±(1.2% of set +1.1% of f.s
.*3
)
Input Voltage Variation
*4
H Range
2mA
4mA
10mA
M Range
2mA
4mA
10mA
L Range
0.1mA
0.2mA
0.6mA
Ripple
RMS
*5
3mA
5mA
20mA
*7
P-P
*6
30mA
50mA
100mA
*7
*1 Full scale of H range
*2 Vin: input terminal voltage of electronic load
*3 M range applies to the full scale of H range
*4 When the input voltage is varied from 1.5V to 150V at a current of rated
power/150V
*5 Measurement frequency bandwidth: 10Hz to 1MHz
*6 Measurement frequency bandwidth: 10Hz to 20MHz
*7 At measurement current of 100A