30

Proper connection of the transmitters with a PC requires an INOR USB Interface (see Figure 17

for connection of the C520/C520X transmitters and figure 18 for connection of the R520/R520X

transmitters).

ATTENTION!
Make sure the area is safe before connecting the INOR USB Interface ICON to the transmitter.

Figure 17: Connection during configuration of C520/C520X

1 Input (intrinsically safe)

2 Output voltage supply (intrinsically safe terminals 6 and 7) and R

Load

(intrinsically safe)

3 Communication with USB Interface and PC software ConSoft

4 Safe area

5 Classified hazardous area (potentially explosive area e.g. Zone 0, 1 or 2)

6 Connection to USB Interface and PC software ConSoft is not permitted if C520X is placed in the

classified hazardous area (potentially explosive area)

7 Connection to the voltage supply during configuration with USB Interface and PC software ConSoft is

not needed

ConSoft

1

2

3

1

2

6

5

4

4