5.5 Diagnostic information according to NAMUR NE 107

Diagnostic information about the connected sensor/sensors and the transmitter is available in

the configuration software ConSoft.

The HART diagnostic messages are generated in accordance with NAMUR NE107.

The following are examples of diagnostic information available both in ConSoft and as HART

messages:
• Sensor break (failure / sensor wire break on Ch1 or Ch2)
• Sensor short-circuit (failure / sensor wire short-circuit on Ch1 or Ch2)
• Sensor low isolation (failure / low internal isolation in sensor on Ch1 or Ch2)
• Sensor drift (failure / sensor drift between Ch1 and Ch2)
• Sensor backup, automatic switchover from defective sensor (maintenance required / change

the sensor on Ch1 or Ch2)

• Transmitter error (failure/transmitter error, non-correctable in the transmitter)

For complete information about all diagnostic information available from the 520 series

transmitters, refer to Table 5, 6, 7 and 8 below.

Messages in AMS, PDM, and FC375 device diagnostic

Actions

Standard

NE107

Signal [12]

Error message

Error correction

Transmitter

C

Primary variable out

of limits

Check the limits for connected sensor. Replace

the sensor allocated to the primary variable.

Transmitter

C

Non-primary variable

out of limits

Check the limits for connected sensor. Replace

the sensor allocated to the non-primary variable.

Transmitter

C

Primary variable

analog output

saturated

Check the limits for connected sensor. Ambient

temperature is higher then stated limits for

sensor, which is allocated to the primary variable

analog output. Replace the sensor allocated to

the primary variable analog output.

Transmitter

I

Primary variable

analog output fixed

Transmitter in multi-drop mode or loop test

Transmitter

C

More status variable

Check device specific errors for more information

Transmitter

I

Cold start

Transmitter is in cold start mode

Transmitter

I

Configuration

changed

Information for other softwares.

Transmitter

C

Field device

malfunction

Make a self test to find out more. Check even

device specific diagnostic. If the error remains

replace the device.

Extended device status

Transmitter

M

Maintenance required Check Device specific diagnostic for more

information about sensors and transmitter.

Transmitter

M

Device variable alert

Check Device specific diagnostic for more

information about sensors and transmitter.

Table 5: The HART status messages and diagnostic information for C520/C520X / R520/R520X transmitters