OPERATION

C520/C520X / R520/R520X

09/2010 • 86B5200001

www.inor.com

35

Messages in AMS, PDM, and FC375

device diagnostic

Actions

Device specific Failure

Error message

Error correction

Transmitter

F Sensor break

channel 1

Sensor wire broken. Check the sensor and sensor connection on the

Channel 1. If error remains replace the sensor connected to Channel 1.

Transmitter

F Sensor break

channel 2

Sensor wire broken. Check the sensor and sensor connection on the

Channel 2. If error remains replace the sensor connected to channel 2.

Transmitter

F Sensor short circuit

channel 1

Sensor wire shorted. Check the sensor and sensor connection on the

Channel 1. If error remains replace the sensor connected to Channel 1.

Transmitter

F Sensor short circuit

channel 2

Sensor wire shorted. Check the sensor and sensor connection on the

Channel 2. If error remains replace the sensor connected to Channel 2.

Transmitter

F Sensor low isolation

channel 1

To low isolation in the sensor. Check the sensor by measuring

resistance between the wires. If error remains replace the sensor

connected to Channel 1.

Transmitter

F Sensor low isolation

channel 2

To low isolation in the sensor. Check the sensor by measuring

resistance between the wires. If error remains replace the sensor

connected to Channel 2.

Transmitter

F Sensor drift

Check the sensors and limits for sensor drift. If error remains replace

the faulty sensor.

Transmitter

F System error

Make a self test to find out more, If the error remains replace the device.

Table 6: The HART status messages and diagnostic information for C520/C520X / R520/R520X transmitters

Messages in AMS, PDM, and FC375

device diagnostic

Actions

Device specific Out of specification

Error message

Error correction

Transmitter

S High limit alarm

OPM

Check the sensor(s) and sensor connection on both channels. If error

remains replace the sensor(s) connected to the transmitter.

Transmitter

S Low limit alarm OPM Check the sensor(s) and sensor connection on both channels. If error

remains replace the sensor(s) connected to the transmitter.

Device specific

Maintenance required
Error message

Error correction

Transmitter

M Sensor backup

Transmitter is measuring the working sensor (channel). Replace the

faulty sensor.

Table 7: The HART status messages and diagnostic information for C520/C520X / R520/R520X transmitters