BM

102

31


Function

Input Range

Description

1.4.0 User data

1.4.3 Checksum

Read only

This value is used for identification of
the device software version. The
checksum is tested when starting.
This helps to detect any problems
with the microcontroller.

1.4.4 Device number**
(Tag number)

00000 01

This parameter assigns an
identification number to the device.
A text consisting of max. 8 ASCII
characters can be entered.

1.4.5 Serial number

Read only

This parameter serves to identify the
respective measuring device. This
number cannot be changed and sets
the address for use with HART

®

interfaces.

1.4.6 French command number Read only

Factory-programmed number, to be
quoted in case of warranty and
service claims.

1.4.7 German command
number

Read only

Factory-programmed number, to be
quoted in case of warranty and
service claims.

1.4.8 Option**
(Descriptor)

This function can be used for entering
text of max. 15 ASCII characters
(customer information only)

Single rod, Twin rod,
Single cable, Single
cable +
counterweight, Single
cable without
counterweight, Twin
cable, Twin cable +
counterweight,
Coaxial, Special 1,
Special 2 or Special 3

1.4.9 Probe type

As per order

Information on probe type supplied
with signal converter. This is a read-
only parameter.


** Refer to section 3.3.5: Characters available for alpha-numerical data functions in PC STAR 2