Fault Clearing PCSTAR 2


Event Fault Action
PCSTAR 2 is unable to
connect to the gauge

The RTS Shutdown parameter
is unable to assign a value to
the PC operating system (XP
for PCSTAR 2 versions prior to
2.01) When at default value 0,
PCSTAR 2 tests the O.S. to
identify which one is installed.
An RTS Shutdown timing value
is assigned depending on the
results. However, this test may
not recognize Windows XP, so
PCSTAR will not work.

Please follow the corrective
procedure below:


Corrective procedure to permit gauge to connect to PCSTAR 2 using OS WINDOWS XP
Step

Action

Data entered/value set

1

Go to PcSt2NT.ini file in the Windows directory on
the hard disk under the PC STAR 2 sub-directory.

n/a

2

Find RTSShutdown in the list of parameters. The
default value is 0.

n/a

3

Modify the line.

RTSShutdown = 150

4 Save

file.

n/a

5

Close the file and run PCSTAR 2. No further
software configuration problems should be
encountered.

n/a