Supplementary procedure for setting the 4 … 20 mA current to output volume readings
(using PCSTAR 2)

Step

Action

Data entered/value selected

1

Go to user function 1.3.1 ”Current 1 item” to select
measurement function.

Volume

2

Go to user function 1.3.2 ”Current 1 range” to set fault
status

4 … 20 mA

3

Go to user function 1.3.3 “Scale I1 min” to enter
volume value for the minimum output (4 mA)

0.50 m³ or 17.66 ft³

4

Go to user function 1.3.3 “Scale I1 max” to enter
volume value for the maximum output (20 mA)

16.80 m³ or 52.97 ft³

5

Either save the new configuration to disk by left-
clicking on F5-Save to disk and download it onto the
BM 102 using the F6-Send to BM102 key.

n/a

6

Press F3 to exit the configuration menu.

n/a



3.3.4 HART

®

Communicator: installation & operating instructions


Display and configuration can also be carried out with a HART

®

communicator. Operator control via

the separate HHC is described in detail in the operating instructions supplied with the HHC.


Basic instructions for taking measurements