KROHNE BM 102 Handbook User Manual

Page 60

Advertising
background image

60

BM 102


Event

Fault

Action

“Flange not found” status
marker on*

The signal converter has been
incorrectly configured to
measure with a cable or rod
probe when it is equipped with
a coaxial probe. This may also
due to installation on a long
nozzle which has the effect of
attenuating the flange pulse.

Contact your local KROHNE
Service Centre for the
corrective procedure**.

“Delay out of limits” status
marker on*, reading is frozen.

The emitted pulse has not
been detected. The gauge will
not work until it has been
found.

The signal converter may
need replacing. Please
contact your local KROHNE
Service Centre**.

“Negative voltage error” status
marker on*
“VC01 voltage error ” status
marker on*
“VC02 voltage error” status
marker on*
“Reprogramming FPGA”
status marker on*

Occurs when there is a
problem with the time base on
the HF board.

Please contact your local
KROHNE Service Centre**.






* Gauge connected to PC-STAR 2 software or HART Handheld Controller (HHC) with Marker

status window (F8) or list displayed

** KROHNE service centre contact details are supplied on the last page of this handbook.