F7 – Oscilloscope function
Using the F7 – Oscilloscope function (Oscillo.), all reflections occurring at the probe can be
displayed. This is a help for optimum setting of the dead zone, the non-measurable zone, and the
threshold under Function F11. It provides details on the position and strength (amplitude) of every
signal between the initial pulse and the tip of the probe.
Refer to section 3.4.2: Gain and signal amplitude for information on how to use F7-Oscilloscope and
F11-Dynamic Configuration windows to modify Gain and level threshold manually to assist detection
of the product surface in unfavourable installation conditions.
Other functions available on the Oscilloscope window:
F1 – Help
On-line help file
F3 – Exit
Exit to real-time measurement main window
F4 – Zoom -
Reduces the time slice scale
F5 – Zoom +
Increases the time slice scale
F10 – Screen copy
To select a directory and save the current screen as a bitmap file.