KROHNE BM 102 Handbook User Manual

Page 81

Advertising
background image

BM

102

81


F7 – Oscilloscope function
Using the F7 – Oscilloscope function (Oscillo.), all reflections occurring at the probe can be
displayed. This is a help for optimum setting of the dead zone, the non-measurable zone, and the
threshold under Function F11. It provides details on the position and strength (amplitude) of every
signal between the initial pulse and the tip of the probe.


Refer to section 3.4.2: Gain and signal amplitude for information on how to use F7-Oscilloscope and
F11-Dynamic Configuration windows to modify Gain and level threshold manually to assist detection
of the product surface in unfavourable installation conditions.

Other functions available on the Oscilloscope window:

F1 – Help

On-line help file

F3 – Exit

Exit to real-time measurement main window

F4 – Zoom -

Reduces the time slice scale

F5 – Zoom +

Increases the time slice scale

F10 – Screen copy

To select a directory and save the current screen as a bitmap file.