KROHNE BM 102 Handbook User Manual

Page 45

Advertising
background image

BM

102

45


Function (Fct.)

Input Range

Description

2.2.0 TESTS

2.2.1 TEST OUTPUT

Select 4 mA, 20
mA, Other

This function allows the current
output to be tested.
The output can be set to one of the
listed values. With a reference
ammeter, the calibration of the
current output can be verified.

2.3.0 SERVICE

Restricted access factory
configuration menu. These
parameters may be accessed via
Fct. 3.2 “Specialist PSW
(password).

3.0 ACCESS

RIGHTS

Yes or no. Enter 9-
character code if
“Yes”.

3.1 MAINTENANCE

PSW

No

Disables the access lock on the
configuration menu.
The password must contain exactly 9
characters.
E, R or U are used only. The
password is displayed in a scrambled
format. It allows KROHNE to decode
the password in case it was forgotten.

3.2

SPECIALIST PSW

See KROHNE
Service centre or
BM 102 Service
Manual for code.

Specialist access to the Service
Parameters.

4.0 WATCH

STATUS

Displays the status of the device.

5.0 HART

®

VARIABLES

5.1 MANUFACTURER

KROHNE

Read only menu.

5.2 MODEL

BM 102

Read only menu.

5.3

FLD DEV REV

1.0

Field device revision.
Read only menu.

5.4 SOFTWARE

REV

1.0

Software revision.
Read only menu.

5.5 HARDWARE

REV

1.0

Hardware revision.
Read only menu.

5.6

DEVICE ID

Read only menu. The device ID is
also the serial number of the device.

5.7

MESSAGE

32 bytes of ASCII characters

5.8

DESCRIPTOR

16 bytes of ASCII characters.
Equivalent to PCSTAR2 Fct. 1.4.8

5.9

DATE

Month Day Year (xx / xx / xx).

5.10 NUM RESP PREAM

Number of preamble in the response
frame of the device.

5.11 TAG

Tag name of the BM 102.

5.12 POLL

ADDRESS

Address of the device.