KROHNE BM 102 Handbook User Manual

Page 33

Advertising
background image

BM

102

33


3.3.3 Quick

Configuration: configuration examples


The minimum functions (fct.) to be configured for a simple measurement are listed below:

Functions Definition
PCSTAR 2

HART

®

1.1.1

2.1.1.1

Tank Height

1.1.2 2.1.1.4 Dead

zone

1.3.1 … 4

2.1.3.1 … 4 Current

Output

1.7.0 2.1.7.0 Volume table*

*For volume measurements

Example procedures for each set of functions are given on the following pages. Each procedure is
given in a series of steps in table form and is according to the PCSTAR 2 F2-Configuration
parameter list. Please refer to section 3.3.4 for the equivalent parameters available on the HART

®

Communicator.

Definitions for quick configuration

* where PCSTAR 2 Fct. 1.3.1 (HART

®

Fct. 2.1.3.1) Current I1 Item is configured to “Level”



Typical gauge used for quick configuration examples:

Probe type:

twin Ø4 mm or 0.15 ” cable probe (type 4)

Product measured:

Water (dielectric constant,

εr = 80)

Tank height:
(PCSTAR 2: Fct. 1.1.1,
HART

®

: Fct. 2.1.1.1)

10000.00 mm or 33 ft.

Dead zone:
(PCSTAR 2: Fct. 1.1.2,
HART

®

: Fct. 2.1.1.4)

0.15 m or ½ ft. (see “probe measurement limits” in section 5.2.3
for the Ø4 mm or 0.15” twin cable probe, type 4)

Probe length, L

1

:

(PCSTAR 2: Fct. 1.1.6
HART

®

: Fct. 2.1.1.2)

9.00 m or 29.5 ft. (do not modify unless advised to)

Advertising