KROHNE BM 102 Handbook User Manual

Page 65

Advertising
background image

BM

102

65


Construction materials

Housing

Aluminium with epoxy coating

Probe

Refer to section 5.2.1: BM 102 mechanical options

Gasket

Viton, optionally Kalrez 6375

Spacers (on twin probes)

Refer to section 5.2.1: BM 102 mechanical options


Weight

Housing

2 kg or 4.4 lb

Single rod Ø 8 mm

0.41 kg/m or 0.28 lb/ft

Single cable Ø 4 mm

0.12 kg/m or 0.08 lb/ft

Twin cable Ø 4 mm

0.24 kg/m or 0.16 lb/ft

Single cable Ø 8 mm

0.41 kg/m or 0.28 lb/ft

Coaxial

1.3 kg/m or 0.87 lb/ft

Standards


The level gauge satisfies the protection requirements of :
Electromagnetic compatibility

Electromagnetic compatibility
directives followed for EU
countries

Directive 89/336/EEC in conjunction with EN 61326-1(A1&A2)
and EN 61000-6-2 standards. It also respects the NAMUR NE
21/98 standard. Protection against EMC emissions is only
guaranteed when the device is used in metal tanks (except when
the device is equipped with coaxial probes).
Low voltage electrical equipment

Other EU directives

Directive 73/23/EEC modified by Directive 93/68/EEC (art.13) in
conjunction with the EN 61010-1 standard.
ATEX*

Directive 94/9/EC in conjunction with EN 50014, EN 50020 and
EN 50284 standards.
Refer to the BM 102 MICROFLEX KEMA 00 ATEX 1101X
Supplementary Installation and Operating Instructions for further
information.


5.1.1 Accuracy
Due to the method of measurement, which primarily determines distance, the measuring accuracy
is specified only as a factor of distance. For that reason, all data in this section refers to the
measured distance.

Standard calibration

Error of measurement (mm)

Distance from flange (m)

Advertising