Specifications – Martel Electronics MasterCAL User Manual

Page 116

Advertising
background image

MasterCAL

Users Manual

106

Fluke 80i-1010 Clamp-on DC/AC Current Probe

Fluke 80i-500s Clamp-on AC Current Probe
(requires the Y9108 adapter)

Fluke 80i-1000s Clamp-on AC Current Probe
(requires the Y91008 adapter)

Fluke 80i-kW Current and Power Probe

Specifications

All specifications apply from +18

°

C to +28

°

C unless

stated otherwise.

All specifications assume a 5 minute warmup period.

Measurement specifications are valid only when
Damping is turned on. When damping is turned off,
or when the

g

annunciator is displayed, floor

specifications are multiplied by 3. Floor specifications
are the second part of the specifications, usually
expressed as "% of full scale." The measure
pressure, temperature, and frequency functions are
specified only with damping on.

The standard specification intervals for the
MasterCAL are 1 and 2 years. Typical 90-day source
and measurement accuracy can be estimated by
dividing the 1 year "% of Reading" or "% of Output"
specifications by 2. Floor specifications, expressed
as "% of f.s.", remain constant.

To achieve the best noise rejection, use battery
power and tie all three common jacks together.