Square-Law Process Variables

When you select

within scaling, the calibrator takes

the square root of its input and displays the
measurement in percent. For example, when
connected to the output of a delta-pressure
transmitter, the calibrator reading is proportional to
flow rate.

1.

If necessary, press

M

for MEASURE mode.

2.

Select a measurement function (

m

,

v

,

h

,

q

,

t

, or

p

) as previously

described.

3.

Press the

Scale

softkey.

4.

Select

scale

from the list.

5.

Use the numeric keypad to enter the 0% of scale
value (

0% Value

).

6.

Press

e

.

7.

Use the numeric keypad to enter the 100% of
scale value (

100% Value

).

8.

Press

e

.

9.

Set

Mode

to

.

10.

Press

e

.

11.

Press the

Done

softkey.

Square root percent of scale remains in effect until
you change to another measurement function or
press the

Scale

softkey and select another scale

mode.