Square-Law Process Variables

When you select

√

within scaling, the calibrator takes

the square root of its input and displays the
measurement in percent. For example, when
connected to the output of a delta-pressure
transmitter, the calibrator reading is proportional to
flow rate.

1.

If necessary, press

M

for MEASURE mode.

2.

Select a measurement function (

m

,

v

,

h

,

q

,

t

, or

p

) as previously

described.

3.

Press the

Scale

softkey.

4.

Select

√

scale

from the list.

5.

Use the numeric keypad to enter the 0% of scale
value (

0% Value

).

6.

Press

e

.

7.

Use the numeric keypad to enter the 100% of
scale value (

100% Value

).

8.

Press

e

.

9.

Set

Mode

to

√

.

10.

Press

e

.

11.

Press the

Done

softkey.

Square root percent of scale remains in effect until
you change to another measurement function or
press the

Scale

softkey and select another scale

mode.