Ramping the output – Martel Electronics MasterCAL User Manual

Page 75

Advertising
background image

Calibrator

Using Source Mode

65

Ramping the Output

Ramping sweeps the source up or down in value.
Use the ramp feature to check a limit switch or
alarm, or any time you want a smoothly increasing or
decreasing output function. You can set the
calibrator to ramp up or down in engineering units
(mV, V, mA, °C, etc.) or % of scale.

During ramping, the output is adjusted 4 times per
second. The size of the steps is determined by your
choices of endpoints and ramp time. For example, if
you set the calibrator to ramp from 1 mV to 1 V over
10 seconds, the output is adjusted in approximately
25 mV steps.

Ramping continues until the selected limit is
reached, or until an optional trip condition is
encountered. The optional trip detect works as
follows: during ramping, the calibrator checks for
either a 1 V change in dc voltage or a change in
continuity status (

Open

or

Short

) from one ¼ second

interval to the next.

Proceed as follows to ramp (i.e., sweep the source):

1.

Refer to the appropriate heading earlier in this
manual (e.g., “Sourcing Electrical Parameters”)
and connect the calibrator to the circuit to be
tested. Figure 22 shows an example.

2.

To automatically stop ramping if a trip condition
is detected, connect a voltage trip circuit to the V
MEAS jacks or a continuity trip circuit to the mA

RTD MEAS jacks. (Continuity detection is not

available when sourcing current.)

3.

If necessary, press

M

for SOURCE mode.

4.

Set the calibrator for the desired source value as
previously described.

5.

If you want to ramp the output in % of scale, set
% of scale as described previously under
“Sourcing in Percent of Scale.”

6.

Press the

More Choices

softkey.

Advertising