Dc voltage measurement – Martel Electronics MasterCAL User Manual

Page 117

Advertising
background image

Calibrator

Specifications

107

DC Voltage Measurement

Range

Resolution

% of Reading +% of Full Scale

1 Year

2 Year

110 mV

1

µ

V

0.025% + 0.015%

0.05% + 0.015%

1.1 V

10

µ

V

0.025% + 0.005%

0.05% + 0.005%

11 V

100

µ

V

0.025% + 0.005%

0.05% + 0.005%

110 V

1 mV

0.05% + 0.005%

0.1% + 0.005%

300 V

10 mV

0.05% + 0.005%

0.1% + 0.005%

Temperature Coefficient: (0.001% of rdg. + 0.0015% f.s.)/

°

C in the ranges -10 to 18

°

C and 28 to 50

°

C

Input Impedance: 5 M

Common Mode Error: 0.008% f.s./(Common Mode Volt)

Maximum Input Voltage: 300 V rms

Advertising