Using Auto Step

To have the calibrator make a sequence of steps
automatically, either once through the sequence or
repetitively, proceed as follows:

1.

Refer to the appropriate “Using Source Mode”
subheading earlier in this manual (e.g.,
“Sourcing Electrical Parameters”) and connect
the calibrator to the circuit to be tested.

2.

If necessary, press

M

for SOURCE mode.

3.

Set the calibrator for the desired source value.

4.

If you want to step the source value in % of
scale, set the % of scale value as described
previously under “Sourcing in Percent of Scale.”

5.

Press the

Step

softkey.

6.

Press the

Auto Step

softkey.

7.

Select values for the following parameters as
you are prompted by the display:

Starting point (in units or % of scale)

Ending point

Number of steps

Time per step

Single shot or continuous repetition

Ramp pattern if continuous

Start delay

8.

To start automatic stepping, press the

Start Step

softkey. The softkey label changes to

Stop Step

.

9.

To stop automatic stepping, press the

Stop Step

softkey.

10.

Press the

Done

softkey to resume normal

operation.