Testing continuity, Measuring pressure – Martel Electronics MasterCAL User Manual

Page 40

Advertising
background image

MasterCAL

Users Manual

30

Note

When measuring frequency, you are
prompted to select a frequency range. If you
expect the frequency you are measuring to
be below 20 Hz, press

d

to select the

lower frequency range, then press

e

.

2.

Connect the test leads as shown in Figure 10,
depending on the measurement function.

Testing Continuity

When testing continuity, the beeper sounds and the
word

Short

appears on the display when the

resistance between the

MEAS jack and its

common jack is less than 25

. The word

Open

appears when the resistance is greater than 400

.

Proceed as follows to test continuity:

1.

Remove power from the circuit to be tested.

2.

If necessary, press

M

for MEASURE mode.

3.

Press

q

twice so that

Open

appears.

4.

Connect the calibrator to the circuit to be tested
as Figure 10 shows.

Measuring Pressure

Many ranges and types of pressure modules are
available from Beta. See “Accessories” near the
back of this manual. Before you use a pressure
module, read its Instruction Sheet. The modules
vary in how you use them, how you zero them, what
types of process pressure media are allowed, and
accuracy specification.

To measure pressure, attach the appropriate
pressure module for the process pressure to be
tested as described in the module’s Instruction
Sheet.

Proceed as follows to measure pressure:

w

Warning

To avoid a violent release of pressure in
a pressurized system, shut off the valve
and slowly bleed off the pressure before
you attach the pressure module to the
pressure line.